    Determination of 6 Penicillin Residues in Honey——The Choice of Analysis Conditions

    • Last Update: 2021-09-20
    • Source: Internet
    • Author: User
    Search more information of high quality chemicals, good prices and reliable suppliers, visit www.echemi.com

    5.
    2.
    4.
    7 Selection of analysis conditions

    (1) Selection of solid phase extraction column for purification sample

    In the study of this method, Oasis HLB solid phase extraction column, BUND ELUT C 18 solid phase extraction column, ENVI-18 solid phase extraction column, Octadecyl C 18 solid phase extraction column, LiChrolut EN solid phase extraction column and Sep-Pak were used The C 18 solid phase extraction column has a total of six solid phase extraction columns to compare the extraction efficiency and purification effects of six penicillins
    .


    Experiments have found that the Sep-Pak C18 solid phase extraction column has a low recovery rate (only 50% to 60%) and the purification effect is not ideal.


    (2) Selection of extraction solution

    Deionized water, 4% sodium chloride solution and 0.
    1 mol/L sodium dihydrogen phosphate buffer solution (pH 9.
    0) were used as extracting solutions to complete the recovery rate experiment of 6 penicillins in honey samples
    .


    The experiment found that the recovery rates of the three extracts were in the range of 78.


    Table 5-29 The influence of three extracts on recovery rate (%)

    (3) Selection of eluent and dosage of OasisHLB solid phase extraction column

    This method compares the effects of using acetonitrile and methanol as the eluent of OasisHLB solid phase extraction column on the recovery of six penicillins
    .


    The experiment found that using acetonitrile as the eluent and the amount within 3 mL, all 6 penicillins could be completely eluted from the OasisHLB column


    The elution conditions of 6 penicillins were tested with 0.
    5-5 mL methanol respectively, and the results are shown in Table 5-30
    .


    From Table 5-30, it can be seen that the recovery rate of OasisHLB solid phase extraction column with 3mL methanol eluting 6 kinds of penicillins is above 90%.


    Table 5-30 The influence of the amount of eluent of OasisHLB solid phase extraction column on the recovery rate (%)

    (4) Selection of liquid chromatography analysis column

    The μBondapak C 18 300mm×4.
    0mm chromatographic column and the SunFire TM 3.
    5um C 18 150mm×2.
    1mm chromatographic column were compared respectively
    .


    Experiments have found that the sensitivity of these two chromatographic columns can meet the requirements, but the μBondapak C 18 300mm×4.


     

     

    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.