Figure 1 Sample import
Figure 2 Sample injection
4.
4.
5 Reagents and materials
5.
5.
Note: Compressed ammonia or nitrogen is a high-pressure gas,
5.
Note: Hydrogen is an extremely flammable high-pressure gas
5.
6 sampling
According to the technical requirements of GB/T 13289, if the liquid sample is taken, the corresponding gasification device should be used
7 Instrument preparation
7.
Column temperature: initial temperature: 35℃; initial temperature retention time: 2.
Split injection system: injection valve loop volume: 200uL~500uL, injection valve temperature: 35℃~45℃, split injector temperature: 150℃~200℃, split ratio: 50:1~100:1, FID temperature: 300°C, air: 300 mL/min, oxygen: 30 mL/min, makeup gas N 2 , 20 mL/mnin
Sensitivity: Set at an appropriate value that can obtain the impurity measurement value
Splitless injection system: injection valve loop volume: 10uL~60uL, injection valve temperature: 35℃~45℃
Note: The heating of 1 Al 2 O 3 PLOT column should not exceed 200°C to prevent changes in column activity
2 The air and hydrogen flow parameters of FID can be determined according to the values recommended by the instrument manufacturer
.
7.
2 When the instrument is stable, it can be analyzed
.