    1 Scope

    This standard specifies the method for determining the residual internal acrylonitrile monomer content in styrene - acrylonitrile copolymers and blends by gas chromatography .


    2 Normative application documents

    The clauses in the following documents have become clauses of this standard after being quoted in this standard


    ISO2561 Plastics-Determination of Residual Styrene Monomers in Polystyrene and Impact Polystyrene-Gas Chromatography

    3 Principle

    Dissolve the sample with dimethylformamide.


    4 Reagents

    During the test, only use approved analytical reagents.


    4.


    4.


    4.


    5 Instruments

    5.


    Test conditions of gas chromatograph:

    Chromatography: It is recommended to use stainless steel or glass tubes with a length of 1m~2m and an inner diameter of 3mm~4mm, filled with PorapakQ with a particle size of 150um~270um (that is, 50 mesh to 100 mesh)


    The filling method of the chromatographic rod is not specified, but it is required to obtain a satisfactory separation effect.


    Temperature of chromatographic column (including preparatory column if used): 160℃~180℃


    Inlet temperature: 200℃~230℃


    Detector temperature: 230℃


    Carrier gas: Ammonia (or nitrogen), which meets the requirements of ISO2561


    Adjust the carrier gas flow rate so that the peak time of propionitrile is 5min to 10min


    Adjust the oxygen and air flow rates of the flame ionization detector to obtain the following results:

    a) High sensitivity response:

    b) There is a linear response in the measured range;

    c) Small changes in gas flow rate have little effect on response or sensitivity


    There should be a layer of polytetrafluoroethylene on each side of the silica gel pad used for sample injection to prevent the corrosion of dimethylformamide
    .
    If the rubber pad has only one-sided protective layer, two layers should be used.
    Even so, the rubber pad must be replaced frequently.

    Note: Capillary columns that can achieve the same separation effect, such as FFAP, can be used
    .

    5.
    2 Micro syringe: the volume is 1uL to 50uL
    .

    5.
    3 Analytical balance: accuracy 0.
    5mg
    .

    6 Preparation of samples

    Samples can be taken from powdered materials, pellets or molded articles
    .
    Large pieces of material should be crushed into small enough fragments to be able to weigh a sample as close as possible to 1.
    0g.
    The sample should not be heated during the crushing process
    .

    Related Links: Industrial Acrylonitrile—Determination of Copper Content Graphite Furnace Atomic Absorption Method (2)

     

     

    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.

    Contact Us

    The source of this page with content of products and services is from Internet, which doesn't represent ECHEMI's opinion. If you have any queries, please write to service@echemi.com. It will be replied within 5 days.

    Moreover, if you find any instances of plagiarism from the page, please send email to service@echemi.com with relevant evidence.