    1 Scope

    This standard specifies the test methods for the determination of trace moisture in fluoroalkanes for industrial use: Karl Fischer method-one-volume method, Karl Fischer method-coulometric method and electrolysis method


    This standard applies to the determination of trace moisture in industrial fluoroalkanes

    2 Normative references

    The clauses in the following documents have become the clauses of the wooden standard through the reference of this standard


    GB/T6283-1986 Determination of water content in chemical products Karl Fischer method (general method)

    GB/T6682-1992 Analysis laboratory water specifications and test methods (GB/T6682-1992, neqISO3696: 1987)

    3 general rules

    3.


    3.


    3.


    3.


    4 Sampling and sample pretreatment·

    According to the corresponding product standard regulations


    5 Test method

    5.


    Some test procedures specified by the test method may lead to dangerous situations


    5.


    5.


    The water in the sample reacts quantitatively with Karl Fischer reagent, and the Karl Fischer reagent is calibrated accurately with water in advance, and the end point is judged by the "dead stop"


    Reaction formula: H 2 O+I 2 +SO 2 +3C 5 H 5 N+CH 3 OH=2C 5 H 5 N·HI+C 5 H 5 NH·SO 4 CH 3 (light brownish yellow)

    5.


    5.


    Figure 1 Schematic diagram of sampling cylinder

    5.


    Figure 2 Schematic diagram of trace moisture analyzer

    5.


    5.
    2.
    2.
    4 Sampling needle: the length of the needle is 150mm~200mm, the inner diameter is 0.
    5mm~0.
    7mm, the schematic diagram is shown in Figure 3
    .

    Figure 3 Schematic diagram of the injection needle

    5.
    2.
    2.
    5 Balance; the maximum weighing is not less than 3000g, and the division value is 0.
    1g
    .

    5.
    2.
    2.
    6 Electronic balance: the graduation value is 0.
    0001g
    .

    5.
    2.
    2.
    7 Micro syringe: 10uL
    .

    5.
    2.
    2.
    8 Glass syringe: not less than 5mL
    .

    5.
    2.
    2.
    9 Micro burette: 5mL
    .

    Related link: Measured value of relative mass correction factor of each component of fluoromethane

     

     

    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.