    Dialysis blood chamber fit tester adopts 7-inch monochrome LCD touch screen display and Chinese menu display
    .
    And by the keyboard to select the nominal capacity of the product, display mode: can arbitrarily set the pressure output value, can print out the report, display time adjustable Executive standard: YY 0053-2016GB/T1962.
    2 syringe injection needle and other medical devices 6% (Luer) conical joint Part 2: locking joint (GB/T1962.
    2-2001; ISO 594-2:1998, IDT) GB/T14233.
    1-2008 medical infusion, blood transfusion, Test methods for injection devices Part 1: Chemical analysis methods GB/T16886.
    1 Biological evaluation of medical devices Part 1: Evaluation and test in the process of risk management (GB/T16886.
    1-2011; ISO10993-1:2009, IDT) YY 0053-2016, YY 0267-2016, GB8368-2018
    。 Number of technical productions; 1, the instrument output pressure is required to be stable, no leakage, pressure, ) can reach -100kPa~500kpa, positive pressure and negative pressure can be measured; 2, with waterproof function, with waterproof filter, supporting customized Luer joint, dialysis joint; 3, operation interface: Chinese Simplified/English, seven-inch color touch screen 4, running time: 0~9999 (settable) 5, PLC control system; 6, color 7-inch touch screen, Chinese and English switching; 7.
    One set of test fixture; 8.
    1 set of test software
    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.