: FICOLL after centrifugal interface is not clear what is the general reason?
: There are many reasons:
1, FICOLL does not have room temperature balance;
2, blood dilution is not fully mixed;
3, when the blood is not layered clearly;
4, blood anticoagulant is not sufficient, cell block, it is recommended to take good blood after room temperature for half an hour to 1 hour period from time to time mixed;
5, individual differences, the same experimental conditions, different sources of blood may vary;
6, centrifugal speed is too small to have been completely stratified, or during centrifuge speed instability;
7, speed reduction too fast, speed reduction time should be more than 5min.
: 1, the proportion should be appropriate: diluted blood: lymphocyte separation fluid is 1:1.
2, blood tiled to lymphocyte separation fluid is clearly layered?
3,
centrifuges can not
brake suddenly, should let the speed slowly down.
.
This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only.
This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of
the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed
description of the concern or complaint, to
service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content
will be removed immediately.