Glucose oxidase method to determine blood sugar content.
-
Last Update: 2020-10-20
-
Source: Internet
-
Author: User
Search more information of high quality chemicals, good prices and reliable suppliers, visit
www.echemi.com
"Purposes and Requirements"learn and master the method of
glucose
oxidase method to determine blood sugar content.The
Of Experiments see Trinder. P. Ann. Clin. Biochem., 1969, 6:24-27.quinoa has the maximum light absorption peak in the range of 480-550nm, and the amount of amine is directly related to the amount of glucose, the amount of niacin can be calculated by using color method to determine the amount of nicotinamide.
Reagents
and Equipment)(i) Reagents
Reagent Stability: Original reagents are stored in 2 to 8 degrees C, valid for 15 months. After mixing, 2 to 8 degrees C can be stabilized for 1 month, room temperature can be stabilized for 3 days.
samples: fresh insoluble
serum
.
(ii) Equipment: . Test tube
and test tube rack, straw, pipe, piped gun,
stration temperature
water bath, 722 type
hydrometer
.
:
wavelength: 505nm (480 to 550nm) Temperature: 37
C; Mix 10mlR1 with 90mlR2, both working fluids. Table operation below:
2. Mix evenly, 37 degrees C insulation 10 to 15 minutes (avoid direct sunlight), with blank tube zero, respectively, read A standard and A sample.
note is not . 1. The glucose concentration in the sample exceeds 22.2 mmol/L (400 mg/dl), the sample volume can be halved to 5ul, the standard amount remains unchanged, retested, the result multiplied by 2.
2. Reagent R1 is a liquid enzyme, pay attention to antifreeze .
This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only.
This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of
the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed
description of the concern or complaint, to
service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content
will be removed immediately.