4 Test method
The reagents and water used in this standard are all analytical reagents and tertiary water specified in GB/T6682
The standard titration solutions, preparations and products required in the test shall be prepared in accordance with GB/T601 and GB/T603 unless other requirements are specified
4.
Take about 10 mL of the sample cooled to the degree of non-boiling in a test tube with an inner diameter of about 15 mm
4.
4.
To dibutyl phthalate (DBP) and polyethylene glycol 2000 (PEG-2000) in methylene chloride mixed stationary phase separation of the components, detected by a thermal conductivity detector, to calculate the area normalization chloride content
4.
4.
4.
4.
4.
4.
4.
4.
Figure 1 Schematic diagram of the sampling device process
The pore size of the stainless steel sintered sand core in the metal filter is 2 μm to 4 μm to filter out possible mechanical impurities in the sample and protect the injection valve
4.
2.
3.
3 Sampler: 250mL~1000mL stainless steel cylinder with polished inner wall.
The structure size is shown in Figure 2
.
Figure 2 Sampler
4.
2.
3.
4 The structure of the sampler pipe is shown in Figure 3
.
Figure 3 Sampler catheter
4.
2.
3.
5 Chromatographic column: a stainless steel packed column with a length of 3m and an inner diameter of 3mm~4mm
.
4.
2.
3.
6 Chromatographic data processor or chromatographic workstation
.