The impurity of methylamine aqueous solution was determined by gas chromatography , detected by thermal conductivity detector, and quantified by external standard method
The minimum detectable concentration of impurities in each amine is shown in Table 1
Table 1 The lowest detectable concentration of impurities in industrial methylamine aqueous solution
3.
3.
3.
3.
3.
3.
3.
Equipped with a thermal conductivity detector, the temperature accuracy of the thermostat is ±0.
3.
Column tube: a stainless steel column with a column length of 2m and an inner diameter of 2mm or 3mm, or a column length and inner diameter that can meet the separation requirements selected by the user
Weigh about 0.
Packing of the chromatographic column: after the outlet end of the chromatographic column (the end with the detector) is blocked with a little glass wool and gauze, vacuumize, tap the chromatographic column lightly with a small wooden pole, and slowly pour the estimated phase until the stationary phase Can not be installed slowly, remove the chromatographic column and plug glass wool on the other end
.
The packing amount of the stationary phase is about 2.
2g for a column with an inner diameter of 2mm and about 4.
8g for a column with an inner diameter of 3mm
.
Aging of the chromatographic column: The newly prepared chromatographic column should be aged for at least 8 hours under a carrier gas (nitrogen) with a column temperature of 100°C and a flow of about 30 mL/min.
The end of the column should be separated from the detector during aging
.
3.
2.
3.
3.
3 Chromatographic data processor or recorder
.
3.
2.
3.
3.
4 Injector: 10uL micro syringe
.
3.
2.
3.
4 Standard sample
The standard sample of methylamine aqueous solution containing each impurity component is prepared by weighing method, and its concentration is close to the concentration of the tested sample
.
The pure sample requirements for preparing the standard sample are: monomethylamine , dimethylamine , trimethylamine , ammonia solution samples are injected 2uL under the following instrument operating conditions without impurity peaks
.
When there is no pure sample, the impurity blank can be reduced
.