    Operation flow of infrared spectrophotometer

    • Last Update: 2015-03-11
    • Source: Internet
    • Author: User
    Search more information of high quality chemicals, good prices and reliable suppliers, visit www.echemi.com
    Operation process of infrared spectrophotometer 1 Turn on the computer to enter the operating system 2 Turn on the power supply of the infrared host, and double-click the infrared icon on the desktop after the two green indicator lights flash out (* to install or replace the USB port of the computer host, you need to install the driver software *The next time you install the computer, you will find new and unknown equipment Put the CD-ROM provided by the instrument into the CD-ROM drive, and click "next" continuously If the installation is correct, the USB device with "company name" will be found in "universal serial bus controller" in "device manager", which means the connection is normal) 3 The main engine starts to initialize, carry out filter, slit, grating, correction self inspection 4 After the self check is passed, observe whether the real-time monitoring data position transmittance is near 100%, if not, or fluctuates greatly Check whether there is sample rack and sample in the sample bin If there is sample rack and sample removed, click "correct" to recalibrate 5 Place the pressed potassium bromide tablet in the reference window, and place the tablet with sample in the sample bin 6 Observe the real-time monitoring data, check the transmittance, and compare with the infrared spectrum The transmittance position should be appropriate, otherwise the tablet will be recompressed 7 Click "one way" to start scanning 8 After scanning, compare with infrared spectrum 9 Click "automatic peak searching" to mark the position and transmittance of each peak (it can be omitted according to requirements) 10 Fill in relevant information in "test environment" and "register information" (it can be omitted as required) 11 Select the paper direction as "horizontal" in "print setting" 12 Click "print preview" to check whether the result is correct, and then click "print" 13 Click "save" to save the atlas to the folder you created 14 Take out all samples, click "reset", and then turn off the power of the host.
    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.

    Contact Us

    The source of this page with content of products and services is from Internet, which doesn't represent ECHEMI's opinion. If you have any queries, please write to service@echemi.com. It will be replied within 5 days.

    Moreover, if you find any instances of plagiarism from the page, please send email to service@echemi.com with relevant evidence.