echemi logo
Product
  • Product
  • Supplier
  • Inquiry
    Home > Chemicals Industry > Chemical Technology > Solid waste-determination of polycyclic aromatic hydrocarbons-liquid chromatography (2)

    Solid waste-determination of polycyclic aromatic hydrocarbons-liquid chromatography (2)

    • Last Update: 2022-03-10
    • Source: Internet
    • Author: User
    Search more information of high quality chemicals, good prices and reliable suppliers, visit www.echemi.com

    Six, analysis and testing

    (1) Instrument reference conditions (for reference only, can be adjusted appropriately according to the actual instrument)

    Mobile phase: Solution A is acetonitrile, and solvent B is ultrapure water


    Flow rate: 1.


    Injection volume: 10uL


    Column temperature: 30℃


    The gradient changes are shown in Table 3-8


    Table 3-8 Gradient changes of mobile phase of PAH special column


    Detector: Diode Array Detector (PDA), detection wavelength: 228nm


    Fluorescence detector (RF), the detection wavelength is shown in Table 3-9


    Table 3-9 Wavelength gradient change of fluorescence detector


    (2) Calibration curve

    Prepare a standard series of more than 5 points (such as 2.


    Table 3-10 Retention time


    7.


    (1) Qualitative target compound

    Qualitatively based on the retention time of each component of the standard material


    (2) Quantitative calculation

    1.


    See Solid Waste—Determination of Organophosphorus Pesticides—Gas Chromatography (1) for the calculation of solid waste leachate


    2.


    See Solid Waste—Determination of Organophosphorus Pesticide—Gas Chromatography (1) for calculation of solid and semi-solid waste
    .

    8.
    Quality Assurance and Quality Control

    (1) The analysis result of the blank test should be less than the detection limit of the method
    .

    (2) At least one full program blank for each batch of samples (not more than 20 samples per batch)
    .
    The analysis result should meet the control index of the blank test
    .

    (3) The analysis of each batch of samples needs to establish a calibration curve, the correlation coefficient is not less than 0.
    995, the intermediate mass concentration standard solution of the calibration curve should be measured every 20 samples, and the absolute value of the relative error between the measurement result and the standard value should not exceed 10% , Otherwise you need to find out the reason or redraw the calibration curve
    .

    (4) The preparation of the standard series should be at least 5 points, and the preparation mass concentration range should cover the mass concentration of the sample to be tested
    .

    (5) At least one parallel analysis and matrix spike analysis should be done for each batch of samples (not more than 20 samples per batch)
    .
    The relative deviation of the determination results of two parallel samples should be within 30%; the recovery rate of each component in the spiked sample should be between 60% and 120%
    .

    Nine, matters needing attention

    (1) The sampling amount of solid waste samples is generally recommended to be 2-10g, and the specific sampling amount depends on the concentration of the target substance and the complexity of the matrix;

    (2) If the processed sample cannot be analyzed in time, it should be stored in a refrigerated, dark, and air-tight manner at 4°C, and the analysis should be completed within 30 days;

    (3) When performing ASE extraction of solid waste samples, appropriate amount of diatomaceous earth should be added and mixed well to disperse well, otherwise the sample extraction will be incomplete;

    (4) During rotary evaporation and concentration, pay attention to the proper pressure and temperature.
    Too low pressure or too high temperature may reduce the recovery rate of volatile polycyclic aromatic hydrocarbons:

    (5) Transfer after concentration by rotary evaporation, pay attention to washing the pear-shaped bottle several times with a small amount of solvent, and transfer them together, otherwise the target compound will be lost;

    (6) Polycyclic aromatic hydrocarbon compounds are widely distributed in ambient air, water and soil, and the highest content is naphthalene.
    The materials used in the experiment must be processed accordingly to control the blank of the whole procedure
    .

    (7) Difluorobiphenyl can be referred to as a substitute, but under the above chromatographic conditions, a well-known brand of difluorobiphenyl standard solution has two chromatographic peaks.
    If the chromatographic peak is selected at 21 min, the recovery rate is 70% to 120%
    .

     

     

    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.

    Contact Us

    The source of this page with content of products and services is from Internet, which doesn't represent ECHEMI's opinion. If you have any queries, please write to service@echemi.com. It will be replied within 5 days.

    Moreover, if you find any instances of plagiarism from the page, please send email to service@echemi.com with relevant evidence.