    Six, analysis and testing

    (1) Reference conditions of gas chromatography (for reference only, can be adjusted appropriately according to the actual instrument)

    Temperature rotation of the injection port: 280℃


    Carrier gas: nitrogen, purity ≥99.


    The chromatographic column connected to the forward inlet is an auxiliary qualitative chromatographic column, and the other conditions of the qualitative analysis are the same as those of the forward inlet analysis


    (2) Calibration curve

    The dilution solvent is n-hexane, and the standard series (such as 0.


    Table 3-3 Toxaphene quantitative peak retention time


    7.


    (1) Qualitative target compound

    Qualitatively based on the retention time of each component of the standard material


    (2) Quantitative calculation

    The target compound was quantified by an external standard method, and the mass concentration of the target compound in the leachate ρ(uL) was calculated according to formula (1)


    ρ=ρ x ×(V x /V i ) (1)

    Where ρ x — the mass concentration of the target compound calculated from the calibration curve, ug/L;

    V x —Concentrated constant volume of the extract, mL;

    V i — sampling volume of leaching solution, mL


    8.


    (1) Use the linear fitting curve to calibrate, the correlation coefficient should be greater than or equal to 0.


    (2) Before analyzing each batch of samples (up to 10 samples), check the calibration curve with the mass concentration of the middle point of the calibration curve.


    (3) Each batch of samples (up to 10 samples) should be a laboratory blank, and the mass concentration of the target compound in the experimental blank should be less than the detection limit of the method


    (4) Each batch of samples (up to 10 samples) should be subjected to at least one parallel determination, and the relative deviation of the parallel sample determination results should be within 30%


    (5) Each batch of samples (up to 10 samples) should be spiked with actual samples at least once


    Nine, matters needing attention

    (1) The solvents and reagents used in the experiment have a certain degree of toxicity, so the sample pretreatment process should be carried out in a fume hood, and the operator should do his own protection work


    (2) The rotary evaporation should not be too fast and the vacuum degree should not be too low, otherwise it will affect the recovery rate of toxaphene
    .

    (3) The sampling amount of the leaching solution can be appropriately reduced according to the content of the target substance.
    The leaching solution should be extracted immediately, otherwise it should be stored in a refrigerator at 4°C
    .

    (4) The emulsification of the extract can be broken by mechanical separation or centrifugation
    .

    (5) Double columns should be used for qualitative analysis of detected samples, and the accuracy of the results should be judged based on the recovery rate of standard addition
    .

     

     

    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.