echemi logo
Product
  • Product
  • Supplier
  • Inquiry
    Home > Food News > Nutrition News > Summary of failures and troubleshooting methods of capillary electrophoresis

    Summary of failures and troubleshooting methods of capillary electrophoresis

    • Last Update: 2021-11-01
    • Source: Internet
    • Author: User
    Search more information of high quality chemicals, good prices and reliable suppliers, visit www.echemi.com

    1.


    1.


    Possible cause: The capillary is blocked or broken


    Solution: Rinse the capillary with water and observe if there is water flowing out


    2.


    Possible cause: bubbles are generated in the buffer or the sample is precipitated in the zone


    Solution: Degas the buffer ultrasonically.


    3.


    Possible cause: The sample injection volume is too large


    Solution: Reduce the injection volume


    4.


    Possible cause: The sample concentration is too low


    Solution: Use a high-concentration sample to test.


    5.


    Possible cause: The detection wavelength is set incorrectly


    Solution: Confirm the characteristic absorption of the analyte and check the detection wavelength setting


    6.
    Inspection result: the current is normal
    .

    Possible cause: Wrong separation polarity
    .

    Solution: For protein samples, pay attention to the PI and charge of the protein under the separation conditions
    .
    For nucleic acid samples, they are negatively charged under normal conditions
    .

    7.
    Inspection result: the current is normal
    .

    Possible cause: The sample is adsorbed on the inner wall of the capillary
    .

    Solution: For protein and nucleic acid samples, try to use coated capillaries, extreme pH conditions or dynamic coatings to prevent sample adsorption
    .

    8.
    Inspection result: the current is normal
    .

    Possible cause: The optical detector or optical fiber is damaged
    .

    Solution: Perform a standard sample test.
    If there is no corresponding result, there may be a hardware problem
    .

    Two, electrophoresis peak tailing

    1.
    Possible cause: The capillary window is not open or the window position is not correct
    .

    Solution: Open the capillary detection window or adjust the position of the window
    .

    2.
    Possible cause: The sample is adsorbed on the inner wall of the capillary
    .

    Solution: For protein and nucleic acid samples, try to use coated capillaries, extreme pH conditions or dynamic coatings to prevent sample adsorption
    .

    Third, the electrophoresis peak is asymmetry

    1.
    Possible cause: The incision of the capillary inlet is not flat
    .

    Solution: Re-cut the capillary inlet
    .

    2.
    Possible cause: The electrolysis rate of the buffer solution and the sample solution is too different
    .

    Solution: Use buffer as sample solvent
    .

    Fourth, the electrophoresis peak is too wide

    1.
    Inspection results: reduce the sample injection volume and improve
    .

    Possible cause: The sample concentration is too high
    .

    Solution: Reduce the sample concentration or reduce the injection volume
    .

    2.
    Inspection results: reduce the sample injection volume, no improvement
    .

    Possible reason: the sample itself is not uniform in nature
    .

    Solution: This reason is mainly for protein samples, and the probability of occurrence of small molecule samples is low
    .

    5.
    Unstable migration time

    1.
    Inspection result: the peak time is unstable and irregular
    .

    Possible cause: The buffer solution balances slowly with the inner wall of the capillary
    .

    Solution: Avoid flushing the capillary with NaOH between each sample run
    .

    2.
    Inspection result: the peak time is delayed in turn
    .

    Possible cause: The substance in the sample is easily adsorbed on the inner wall of the capillary
    .

    Solution: First, rinse the capillary with 0.
    1N NaOH solution for a short time before each sample run, and observe whether the migration time can be repeated
    .
    If the effect is not good, use a coated capillary or process the sample again
    .

    6.
    Unstable peak shape and peak time

    1.
    Check result: Change the type of buffer, the peak shape and peak time are stable
    .

    Possible reasons: the buffer is unstable, easy to electrolyze, or the sample is unstable under the original buffer condition
    .

    Solution: Replace other types of buffers
    .

    2.
    Inspection result: changing the type of buffer, the peak shape and peak time are still unstable
    .

    Possible cause: The substance in the sample is easily adsorbed on the inner wall of the capillary
    .

    Solution: Use a coated capillary or pre-treat the sample
    .

    Seven, current leakage

    1.
    Inspection result: the capillary tube is broken
    .

    Possible cause: The buffer in the capillary flows out, causing current leakage
    .

    Solution: Replace the capillary, and clean the optical fiber head and detection window with water
    .

    2.
    Inspection result: the capillary tube is not broken
    .

    Possible cause: The humidity in the experimental environment is too high
    .

    Solution: Use a dehumidifier or air conditioner
    .
    When the humidity is too high, the desiccant can be placed when the instrument is turned off, but the desiccant must be taken out when the instrument is running
    .

    8.
    Cannot add air pressure

    1.
    Check result: the problem is solved after replacement
    .

    Possible cause: There is liquid at the neck of the bottle, which makes the stopper slippery
    .

    Solution: When filling the bottle with liquid, do not overfill or wet the bottleneck
    .

    2.
    Inspection result: If the bottle stopper has no problem, the vacuum method can be used first, if the vacuum method can be used, but the pressure is still unavailable
    .

    Possible cause: There is a problem with the pressure system
    .

    Solution: overhaul
    .

    Nine, the migration time can be repeated, but the peak area repeatability is not good

    1.
    Inspection result: Re-cut the capillary inlet and the problem is solved
    .

    Possible cause: The capillary inlet incision is not flat
    .

    Solution: Re-cut the capillary inlet
    .

    2.
    Inspection result: Re-cut the capillary inlet, the problem still exists
    .

    Possible cause: If the electric sampling can maintain reproducibility, there is a problem with the pressure control or the gas circuit
    .

    Solution: overhaul
    .

    10.
    The capillary or electrode is easy to break

    1.
    Possible cause: the bottle cap is aging
    .

    Solution: Replace the cap
    .

    2.
    Possible cause: The electrode is not vertical
    .

    Solution: Straighten the electrode
    .

    11.
    Coolant leakage

    1.
    Possible cause: The gasket and the clip in the cartridge are missing or installed in the wrong order
    .

    Solution: Install in strict accordance with the installation order
    .

    2.
    Possible cause: The internal pipeline of the instrument is not tightly sealed
    .

    Solution: overhaul
    .


    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.

    Contact Us

    The source of this page with content of products and services is from Internet, which doesn't represent ECHEMI's opinion. If you have any queries, please write to service@echemi.com. It will be replied within 5 days.

    Moreover, if you find any instances of plagiarism from the page, please send email to service@echemi.com with relevant evidence.