Common fault handling: First, the main power switch cannot be closed, and it is automatically tripped
.
Cause of accident: The fan is stuck and the motor is blocked, or the line has a short circuit
.
Solution: Adjust the position of the fan shaft, or replace the impeller and bearing, and check whether the line is intact
.
Second, the wind speed is low
.
Cause of accident: 1.
Too much
dust accumulation in the primary filter.
2.
High efficiency filter failure
.
Solution: 1.
Clean the primary filter
.
2.
Replace the high efficiency filter
.
Third, the wind speed is low
.
Cause of accident: 1.
The contactor is not working
.
2.
The fan power core has been fused
.
Solution: 1.
Check whether the contact line is normal
.
2.
Replace the melting core
.
Fourth, the fluorescent lamp is not lit
.
Cause of accident: 1.
Lamp or relay damage
.
2.
The lamp power fuse has been fused
.
Solution: Replace the lamp or relay
.