    Determination of trace water in ethylene and propylene for industrial use (2)

    • Last Update: 2022-01-17
    • Source: Internet
    • Author: User
    Search more information of high quality chemicals, good prices and reliable suppliers, visit www.echemi.com

    3.


    3.


    3.


    Note: The anolyte contains an appropriate amount of ethylene glycol (such as 10% of the total volume) to help absorb the trace water in the sample


    3.


    3.


    3.


    3.


    3.


    3.


    Note 1: Other capillary gasification devices and connection methods that can meet the technical requirements of this standard can also be used


    Note 2: When measuring ethylene, the end of the cylinder without pressure gauge should be used as the gas outlet to connect to the capillary tube


    3.


    3.


    Note: It is not necessary to switch the cylinder valve during the propylene determination process, but to control the sample injection by controlling the capillary port in and out of the sealing gasket



    3.
    6 Expression of results

    3.
    6.
    1 Express the moisture content (c 1 ) in the sample by mass concentration (mg/kg ), and calculate it according to formula (1):

    Where:

    m 1 —The absolute value of moisture displayed by the instrument, ug;

    m—the mass of the sample, g
    .

    3.
    6.
    2 Express the water content (c 2 ) in the sample by volume concentration (mL/m³ ), and calculate according to formula (2):

    Where:

    M—The relative molecular mass of ethylene or propylene (28.
    05 for ethylene and 42.
    08 for propylene);

    18.
    01—The relative molecular mass of water
    .

    3.
    6.
    3 Take the arithmetic average of the two repeated determination results as the analysis result, and round it to 0.
    1 mg/kg according to GB/T8170-1987
    .

    3.
    7 Repeatability

    In the same laboratory, by the same operator, using the same instrument and equipment, the same sample with the moisture content in the range of 1mng/kg-50mg/kg is repeated twice.
    The result is obtained under the 95% confidence level.
    The difference should not be more than 3mg/kg
    .

     

     

    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.