    (1) Scope of application

    This method is suitable for the determination of lead and cadmium in solid waste samples and solid waste leachate


    (2) Method principle

    The solid waste sample or solid waste leachate is digested by acid, and then injected into the graphite furnace atomizer.


    (3) Reagents and materials

    (1) Sulfuric acid , nitric acid , hydrochloric acid , hydrofluoric acid , perchloric acid , hydrogen peroxide , and ammonium dihydrogen phosphate are class pure


    (2) Metal lead and metal cadmium are spectrally pure


    (3) Nitric acid solution (1+9), nitric acid solution (1+99), hydrochloric acid solution (1+1)


    (4) Matrix improver: ammonium dihydrogen phosphate solution


    Accurately weigh 13.


    (5) Lead standard stock solution; p=1000mg/L


    Accurately weigh 1.


    (6) Cadmium standard stock solution: p=1000mg/L


    Accurately weigh 1.


    (7) Lead standard intermediate solution: p=100mg/L


    Accurately pipette 10.


    (8) Cadmium standard intermediate solution: p=100mg/L


    Accurately pipette 10.


    (9) Lead standard liquid: p=1.


    Accurately pipette 1.
    00 mL of the lead standard intermediate solution into a 100 mL volumetric flask, dilute to the mark with a nitric acid solution (1+99), and shake well
    .
    Now equipped for temporary use
    .

    (10) Cadmium standard liquid: p=1.
    00mg/L
    .

    Accurately pipette 1.
    00 mL of the cadmium standard intermediate solution into a 100 mL volumetric flask, dilute to the mark with a nitric acid solution (1+99), and shake well
    .
    Now equipped for temporary use
    .

    (4) Instruments and equipment

    (1) Graphite furnace atomic absorption spectrophotometer with background correction function
    .

    Different types of graphite furnace atomic absorption spectrophotometers have different optimal measurement conditions, and the test conditions can be optimized according to the requirements of the instrument manual
    .

    Table 5-8 Reference measurement conditions of the instrument

    (2) Lead hollow cathode lamps, cadmium hollow cathode lamps
    .

    (3) Electric heating plate or graphite digestion instrument: with temperature control function (temperature stable ±5℃), the maximum temperature can be set to 200℃
    .

    (4) Microwave digestion instrument: output power 1000~1600W
    .
    With programmable control function, temperature, pressure and time (heating time and holding time) can be monitored in the whole process; with safety protection function
    .

    (5) Analytical balance: accuracy of 0.
    1mgn

    (6) PTFE crucible: 50mL
    .

    (7) Sieve: non-metallic sieve, 100 mesh
    .

    (8) High-purity argon: the purity is not less than 99.
    999%
    .

    Related Links: Atomic Absorption Spectrometry (3)

     

     

    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.

    Contact Us

    The source of this page with content of products and services is from Internet, which doesn't represent ECHEMI's opinion. If you have any queries, please write to service@echemi.com. It will be replied within 5 days.

    Moreover, if you find any instances of plagiarism from the page, please send email to service@echemi.com with relevant evidence.