echemi logo
Product
  • Product
  • Supplier
  • Inquiry
    Home > Chemicals Industry > Chemical Technology > Solid Waste—Determination of Phenylurea Compounds—Liquid Chromatography (2)

    Solid Waste—Determination of Phenylurea Compounds—Liquid Chromatography (2)

    • Last Update: 2022-03-10
    • Source: Internet
    • Author: User
    Search more information of high quality chemicals, good prices and reliable suppliers, visit www.echemi.com

    Six, analysis and testing

    (1) Instrument conditions (for reference only, can be adjusted appropriately according to the actual instrument)

    Analysis column: C 18 column (250mm×4.


    Mobile phase: solution A is acetonitrile , solvent B is 25mmol/L phosphate buffer


    Flow rate: 1.


    The gradient changes are shown in Table 3-11


    Table 3-11 C18 column mobile phase gradient changes


    Detector: Diode Array Detector (PDA), detection wavelength: 245nm


    Injection volume: 10uL


    Column temperature: 35℃


    In the case of doubtful results, different chromatographic columns can also be used to confirm the results.


    Chromatographic column: cyano column (150mm×4.


    Mobile phase: Phase A is acetonitrile ; Phase B is 25mmol/L phosphate buffer


    Flow rate: 1.


    The gradient changes are shown in Table 3-12


    Table 3-12 Gradient changes of cyano column mobile phase


    Other conditions are the same as those of the above analytical column analysis


    (2) Standard curve

    Prepare a standard series of more than 5 points (such as 2.


    Table 3-13 Retention time

    7.
    Result calculation

    (1) Qualitative target compound

    Qualitatively based on the retention time of each component of the standard material
    .

    (2) Quantitative calculation

    See Solid Waste—Determination of Organophosphorus Pesticides—Gas Chromatography (1) for calculation of solid and semi-solid waste
    .

    8.
    Quality Assurance and Quality Control

    (1) Calibration curve

    Use the linear fitting curve to calibrate, the correlation coefficient should not be less than 0.
    99, otherwise the calibration curve should be redrawn
    .

    (2) Blank

    Each batch of samples (up to 20 samples) should be a laboratory blank, and the mass concentration of the target compound in the experimental blank should be less than the detection limit of the method
    .

    (3) Parallel sample determination

    Each batch of samples (up to 20 samples) should be subjected to at least one parallel determination, and the relative deviation of the parallel sample determination results should be within 20%
    .

    (4) Standard addition of actual samples

    Each batch of samples (up to 10 samples) should be spiked with actual samples at least once, and the spike recovery rate should be between 30% and 140%
    .

    Nine, matters needing attention

    (1) The high-quality concentration reagents and standard solutions used in the experiment are toxic reagents.
    The configuration and use are recommended to be carried out in a fume hood
    .

    (2) Quartz sand should be burned in a muffle furnace at 450°C for 4 hours to remove other impurities before use
    .

    (3) Appropriate amount of diatomaceous earth should be used to mix and disperse the sample to avoid agglomeration of the sample, which is not conducive to extraction
    .

    (4) When evaporating and concentrating, pay attention to the proper pressure.
    Excessive pressure may reduce the recovery rate
    .

    (5) After rinsing the C18 cartridge of the solid cabinet with a methanol-water mixed solution with a volume ratio of 2:8, the water must be fully drained and then eluted.
    If there is residual water, the nitrogen blowing and concentration time will be prolonged, which may cause The recovery rate is reduced
    .

     

     

    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.

    Contact Us

    The source of this page with content of products and services is from Internet, which doesn't represent ECHEMI's opinion. If you have any queries, please write to service@echemi.com. It will be replied within 5 days.

    Moreover, if you find any instances of plagiarism from the page, please send email to service@echemi.com with relevant evidence.