    Circular of the State Food and Drug Administration on the Results of National Medical Device Supervision and Sampling (No. 4) (No. 63 of 2022)

    • Last Update: 2023-02-03
    • Source: Internet
    • Author: User
    Search more information of high quality chemicals, good prices and reliable suppliers, visit www.echemi.com
    In order to strengthen the supervision and management of medical devices and ensure the quality of medical device products, safety and effectiveness, the State Medical Products Administration organized product quality supervision and sampling inspection of 10 varieties such as electric wheelchairs, small steam sterilizers, and uric acid determination kits, and found that 29 batches (sets) of products did not meet the standard provisions
    .
     
    1.
    Medical device products whose items are sampled do not meet the requirements of the standards
     
    (1) 10 electric wheelchairs: Tianjin Taikang Sunshine Technology Development Co.
    , Ltd.
    , Meritz Medical Equipment (Suzhou) Co.
    , Ltd.
    , Mingguang Longway Medical Equipment Technology Co.
    , Ltd.
    , Langfang Taiwang Medical Equipment Co.
    , Ltd.
    , Ningbo Shenyu Medical Equipment Co.
    , Ltd.
    , Zhejiang INOVA Rehabilitation Equipment Co.
    , Ltd.
    , Bazhou Citizen Likang Medical Equipment Co.
    , Ltd.
    production, involving continuous leakage current and patient auxiliary current at normal working temperature, inhibition driving when charging, maximum speed, horizontal road braking, maximum safe slope braking The input power does not meet the standard regulations
    .
     
    (2) 1 small steam sterilizer: produced by Zhiwei (Xiamen) Instrument Co.
    , Ltd.
    , the steam temperature involving equilibrium time, maintenance time and maintenance time does not meet the standard regulations
    .
     
    (3) Uric acid determination kit 1 batch: produced by Ningbo Botai Biotechnology Co.
    , Ltd.
    , the accuracy involved does not meet the standard regulations
    .
     
    (4) 1 X-ray bone densitometer: produced by Hangzhou Yuanxiang Medical Equipment Co.
    , Ltd.
    , the voltage accuracy of the X-ray tube does not meet the standard regulations
    .
     
    (5) 1 dental implant machine: produced by Osstem Implant Co.
    , Ltd.
    , involving no-load speed that does not meet the standard regulations
    .
     
    (6) 6 sets of cold light sources for endoscopes: Shandong Guanlong Medical Supplies Co.
    , Ltd.
    , Nanjing Yanan Special Lighting Electrical Appliance Factory, Shenzhen Enop Medical Equipment Co.
    , Ltd.
    , Beijing Fanxing Optoelectronic Medical Equipment Co.
    , Ltd.
    , PolyDiagnost GmbH (German Boli Medical Technology Co.
    , Ltd.
    ), Xuzhou Jiabao Electronic Technology Co.
    , Ltd.
    production, involving the radiated flux ratio of red, green and blue light, infrared cut-off performance, input power, samples can not be used normally during normal inspection does not meet the standard regulations
     
    (7) Alginate dressing 2 batches: Hangzhou Jiwei Medical Technology Co.
    , Ltd.
    and Robust Medical Supplies Co.
    , Ltd.
    production, involving diffusion that does not meet the standard provisions
    .
     
    (8) 1 contrast agent injection device: produced by Shenzhen Dongda Precision Instrument Technology Development Co.
    , Ltd.
    , the maximum injection pressure involved does not meet the standard regulations
    .
     
    (9) 3 batches of surgical gowns: marked as produced by Xinxiang Hexie Medical Equipment Co.
    , Ltd.
    and Henan Asia-Pacific Medical Supplies Co.
    , Ltd.
    , involving sterility that does not meet the standard provisions
    .
     
    (10) 3 batches of sodium hyaluronate gel for injection: produced by Relais (Beijing) Medical Device Co.
    , Ltd.
    and CG Bio Co.
    , Ltd.
    fine-based biological Co.
    , Ltd.
    , the content of free sodium hyaluronate does not meet the standard provisions
    .
     
    The above sampling does not meet the standard requirements, and the specific conditions of the products are shown in the annex
    .
     
    2.
    For the above-mentioned products that do not meet the standards, the State Medical Products Administration has required the provincial drug regulatory department where the medical device registrant is located to make administrative processing decisions in a timely manner and announce
    them to the public in accordance with the requirements of the Regulations on the Supervision and Administration of Medical Devices, the Measures for the Supervision and Administration of Medical Device Production, and the Administrative Measures for the Recall of Medical Devices 。 The provincial drug regulatory department shall urge the medical device registrant to conduct a risk assessment of the products that do not meet the requirements of the standards, determine the recall level according to the severity of the medical device defect, actively recall the products and disclose the recall information; Supervise and urge enterprises to find out the reasons for unqualified products as soon as possible, formulate corrective measures and put rectification in place
    on schedule.
     
    Notice is hereby given
    .
     
    Attachment: List of products that do not meet the standard requirements for random inspection
    serial number
    Label the product
    name
    Spot checked
    Indicate the medical device registrant and recordation holder
    Specifications and models
    Date of manufacture/batch number/factory number
    sampling
    Inspection
    Non-compliance with standard provisions
    1
    Electric wheelchair car
    Shanxi Kangwanjia Medical Equipment Chain Co.
    , Ltd
    Tianjin Taikang Sunshine Technology Development Co.
    , Ltd
    DYW-459-46A3 
    2021/6/11
    210428
    216
    Shanxi Provincial Drug Administration
    Shaanxi Medical Device Quality Inspection Institute
    Suppress driving while charging
    2
    Electric wheelchair car
    Daqing Dongfeng Pharmacy Chain Co.
    , Ltd.
    Tongcheng Pharmacy
    Tianjin Taikang Sunshine Technology Development Co.
    , Ltd
    DYW-459-46A5 
    2020/1/17
    20 /190705
    Heilongjiang Provincial Drug Administration
    Shaanxi Medical Device Quality Inspection Institute
    Suppress driving while charging
    3
    Electric wheelchair car
    Tianjin Taikang Sunshine Technology Development Co.
    , Ltd
    Tianjin Taikang Sunshine Technology Development Co.
    , Ltd
    DYW-459-46A10 
    2022/1/5
    211205
    001
    Tianjin Drug Administration
    Shaanxi Medical Device Quality Inspection Institute
    Input power; Suppress driving while charging
    4
    Electric wheelchair car
    Meritz Medical Devices (Suzhou) Co.
    , Ltd
    Meritz Medical Devices (Suzhou) Co.
    , Ltd
    P103
    2020-02-23TNX200200005
    Jiangsu Provincial Drug Administration
    Hunan Institute of Drug Inspection and Testing
    Continuous leakage current and patient auxiliary current at normal operating temperature
    5
    Electric wheelchair car
    Tibet Zhanbo Medical Equipment Co.
    , Ltd
    Mingguang Longway Medical Device Technology Co.
    , Ltd
    LW01301A
    December
    6, 2021 20211206
    20211206010
    Tibet Autonomous Region Drug Administration
    Shaanxi Medical Device Quality Inspection Institute
    maximum speed; horizontal road braking; Maximum safe slope braking
    6
    Electric wheelchair car
    Langfang Taiwang Medical Equipment Co.
    , Ltd
    Langfang Taiwang Medical Equipment Co.
    , Ltd
    TW-100W
    20220401 April
    22, 2022
    Hebei Provincial Drug Administration
    Shaanxi Medical Device Quality Inspection Institute
    Continuous leakage current and patient auxiliary current at normal operating temperature; maximum speed; horizontal road braking; Maximum safe slope braking
    7
    Electric wheelchair car
    Ningbo Shenyu Medical Equipment Co.
    , Ltd
    Ningbo Shenyu Medical Equipment Co.
    , Ltd
    W-A805
    2021.
    5
    N0022104
    N0022104075
    Zhejiang Provincial Drug Administration
    Guangdong Institute of Quality Supervision and Inspection of Medical Devices
    Maximum speed
    8
    Electric wheelchair car
    Zhejiang INOVA Rehabilitation Equipment Co.
    , Ltd
    Zhejiang INOVA Rehabilitation Equipment Co.
    , Ltd
    N5909
    March
    02, 2022 B140203909022100
    Zhejiang Provincial Drug Administration
    Guangdong Institute of Quality Supervision and Inspection of Medical Devices
    Input power
    9
    Electric wheelchair car
    Hubei New Special Medicine Pharmacy Chain Co.
    , Ltd
    Zhejiang INOVA Rehabilitation Equipment Co.
    , Ltd
    N5516
    2021 April
    12, 12
    Hubei Provincial Drug Administration
    Guangdong Institute of Quality Supervision and Inspection of Medical Devices
    Input power
    10
    Electric wheelchair car
    Bazhou citizen Likang Medical Equipment Co.
    , Ltd
    Bazhou citizen Likang Medical Equipment Co.
    , Ltd
    BM-100W
    December
    2022, 4 12
    220401
    Hebei Provincial Drug Administration
    Shaanxi Medical Device Quality Inspection Institute
    maximum speed; horizontal road braking; Maximum safe slope braking
    11
    Benchtop automatic pressure steam sterilizer
    Zhiwei (Xiamen) Instrument Co.
    , Ltd
    Zhiwei (Xiamen) Instrument Co.
    , Ltd
    18L/GT18DP 
    2017.
    08.
    07
    E416D003
    Fujian Provincial Drug Administration
    Guangdong Institute of Quality Supervision and Inspection of Medical Devices
    equilibrium time; Duration of maintenance; Steam temperature for holding time
    12
    Uric acid assay kit (enzyme colorimetry)
    Ningbo PATEO Biotechnology Co.
    , Ltd
    Ningbo PATEO Biotechnology Co.
    , Ltd
    Reagent 1: 66 ml×6; Reagent 2: 16 ml×6; Calibrator: 1 ml
    2021 April
    10, 09
    Zhejiang Provincial Drug Administration
    Shanghai Institute of Medical Device Inspection and Research
    accuracy
    13
    Dual-energy X-ray bone densitometer
    Hangzhou Yuanxiang Medical Equipment Co.
    , Ltd
    Hangzhou Yuanxiang Medical Equipment Co.
    , Ltd
    G3
    2022.
    04.
    06
    GMG3220328001
    Zhejiang Provincial Drug Administration
    Liaoning Medical Device Inspection and Testing Institute
    X-ray tube voltage accuracy
    14
    Dental Surgical Devices
    Aoyatai (Beijing) Trading Co.
    , Ltd
    Osstem Implant Co.
    , Ltd.
    SM5
    2021-12-08
    1005567
    Beijing Municipal Drug Administration
    Guangdong Institute of Quality Supervision and Inspection of Medical Devices
    No-load speed
    15
    Medical cold light source
    Shandong Guanlong Medical Supplies Co.
    , Ltd
    Shandong Guanlong Medical Supplies Co.
    , Ltd
    YL-Ⅲ
    2021.
    10.
    19
    21101501
    21016
    Shandong Provincial Drug Administration
    Guangdong Institute of Quality Supervision and Inspection of Medical Devices
    The radiated flux ratio of red, green and blue light
    16
    Medical cold light source
    Nanjing Yanan Special Lighting Electrical Appliance Factory
    Nanjing Yanan Special Lighting Electrical Appliance Factory
    XD-302-250W
    2022 April
    03, 31
    Jiangsu Provincial Drug Administration
    Guangdong Institute of Quality Supervision and Inspection of Medical Devices
    Infrared cut-off performance
    17
    Medical endoscope cold light source
    Shenzhen Inop Medical Equipment Co.
    , Ltd
    Shenzhen Inop Medical Equipment Co.
    , Ltd
    INP-LUM300
    2022-04-12
    INP2022041300102
    Guangdong Provincial Drug Administration
    Zhejiang Medical Device Inspection and Research Institute
    The radiated flux ratio of red, green and blue light
    18
    Medical endoscope cold light source
    Beijing Fanxing Optoelectronics Medical Equipment Co.
    , Ltd
    Beijing Fanxing Optoelectronics Medical Equipment Co.
    , Ltd
    XD-250
    20220411
    X02220201
    Beijing Municipal Drug Administration
    Shandong Medical Device and Drug Packaging Inspection and Research Institute
    Input power
    19
    Cold light source for endoscopes
    Shanghai Blue Line Electronics Co.
    , Ltd
    PolyDiagnost GmbH
    PD-LS-0220
    2021-03-29
    64194755
    Shanghai Municipal Drug Administration
    Zhejiang Medical Device Inspection and Research Institute
    Samples cannot be used properly during normal testing
    20
    LED medical endoscope cold light source
    Xuzhou Jiabao Electronic Technology Co.
    , Ltd
    Xuzhou Jiabao Electronic Technology Co.
    , Ltd
    L50A
    2022.
    3
    2203002
    Jiangsu Provincial Drug Administration
    Guangdong Institute of Quality Supervision and Inspection of Medical Devices
    The radiated flux ratio of red, green and blue light
    21
    Alginate dressings
    Hangzhou Jiwei Medical Technology Co.
    , Ltd
    Hangzhou Jiwei Medical Technology Co.
    , Ltd
    10cm×10cm
    20211227
    CG-001-211201
    Zhejiang Provincial Drug Administration
    Shandong Medical Device and Drug Packaging Inspection and Research Institute
    Diffuse
    22
    Alginate dressings
    Robust Medical Supplies Co.
    , Ltd
    Robust Medical Supplies Co.
    , Ltd
    High gel type 10cm×10cm
    2021-11-28
    20211128
    Guangdong Provincial Drug Administration
    Shandong Medical Device and Drug Packaging Inspection and Research Institute
    Diffuse
    23
    Contrast agent high-pressure syringe
    Shenzhen Dongda Precision Instrument Technology Development Co.
    , Ltd
    Shenzhen Dongda Precision Instrument Technology Development Co.
    , Ltd
    DDI-400C 
    2022-04-15
    D22C2000190
    Guangdong Provincial Drug Administration
    Liaoning Medical Device Inspection and Testing Institute
    Maximum injection pressure
    24
    Single-use surgical gowns
    Kuitun Hospital of Traditional Chinese Medicine of Xinjiang Production and Construction Corps
    Xinxiang Hexie Medical Equipment Co.
    , Ltd
    120cm×120cm 
    2021 April
    10, 8
    Xinjiang Production and Construction Corps Drug Administration Bureau
    Yunnan Medical Device Inspection and Research Institute
    asepsis
    25
    Single-use medical surgical gowns
    Guyuan Aier Eye Hospital
    Henan Asia-Pacific Medical Supplies Co.
    , Ltd
    tuba
    2021 April
    11, 2
    Ningxia Hui Autonomous Region Drug Administration
    Yunnan Medical Device Inspection and Research Institute
    asepsis
    26
    Single-use medical surgical gowns
    The First People's Hospital of Korla City
    Henan Asia-Pacific Medical Supplies Co.
    , Ltd
    tuba
    2021 April
    2, 20
    Xinjiang Uygur Autonomous Region Drug Administration
    Yunnan Medical Device Inspection and Research Institute
    asepsis
    27
    Cross-linked sodium hyaluronate gel for injection
    Relais (Beijing) Medical Equipment Co.
    , Ltd
    Relais (Beijing) Medical Equipment Co.
    , Ltd
    1.
    0mL 
    211119
    211101
    D2435-D2449
    Beijing Municipal Drug Administration
    China Academy of Food and Drug Control
    Free sodium hyaluronate content
    28
    Cross-linked sodium hyaluronate gel for injection
    Shanghai Hongkang Hospital Co.
    , Ltd
    CG Bio Co.
    , Ltd
    Model: DANAE Line Z1 Specification: 1.
    0ml/pcs/box
    2021-04-09
    C1054J1
    Shanghai Municipal Drug Administration
    China Academy of Food and Drug Control
    Free sodium hyaluronate content
    29
    Cross-linked sodium hyaluronate gel for injection
    Beijing Weiner Medical Beauty Clinic
    CG Bio Co.
    , Ltd
    1.
    0ml/pc/box DANAE Line Z1
    2021-03-26
    C1053J1
    Beijing Municipal Drug Administration
    China Academy of Food and Drug Control
    Free sodium hyaluronate content
     
     
    State Food and Drug Administration
     
    January 3, 2023
    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.