The biological decontamination workbench is designed to protect laboratory personnel and research samples from damage, but there are certain requirements
for the laboratory environment and placement.
Therefore, the device needs to be properly installed and placed in a specific area
.
1.
The purification table should not be placed near any common air source in the laboratory, including vents, personnel passages or doors
.
Placing it near one of these sources can cause airflow disruption in biosafety cabinets, increasing the risk
of cross-contamination.
2.
Make sure that the outflow area of the table exhaust filter is not placed under or near
any ceiling exhaust or intake in the room.
The outflow area of the top exhaust filter is very susceptible to destructive airflow, which can affect exhaust and incoming air, disrupt the air balance and lead to unsafe working conditions
.
If the airflow is strong enough, it may exceed the appropriate inflow speed of the biosafety cabinet, which may cause contaminated indoor air to enter the work area
.
3.
It is suitable for biosafety cabinets
with more toxic chemicals that require exhaust hoods.
The placement of the bio-decontamination table is important here, as you need to secure the position of the cabinet so that the exhaust can easily extend to the top
of the exhaust transition on the roof of the equipment.
You also need to make sure that each side of the cabinet has 6 inches of space for maintenance
.
The power outlet connected to the table should also be easy to use, as the power plug can be used as an emergency disconnect device and should not be blocked
.
Maintenance of biological purification workbench: 1, according to the actual use, the initial filter should be removed and cleaned regularly, and the cleaning cycle is generally 3 to 6 months
.
2.
When the ideal cross-sectional wind speed cannot be achieved after normal replacement or cleaning of the primary air filter, the working voltage of the fan should be adjusted to achieve the ideal uniform wind speed, and the working voltage of the fan can be adjusted to 80V~90V
.
3.
After 18 months of general use, when the working voltage of the fan is adjusted to a high point and the ideal wind speed is still not reached, it indicates that there is too much
dust in the high efficiency air filter.
The holes on the filter media have been basically blocked and should be updated
in time.
4.
When replacing the high-efficiency air filter, pay attention to the correct model, specification and size, install the equipment according to the arrow wind direction, and pay attention to the sealing around the filter