    Injection needle tip puncture force tester is a professional for testing the physical properties of the needle tip of the injection needle efficient precision testing instrument, will pay attention to The visha technology friends should know that when using the needle tip puncture force tester injection needle simulated skin is one
    of the necessary accessories.

    Today, Wiscia Technology will explain to you in detail what is the tester program of the injection needle tip puncture force tester? First of all, we review the principle of the injection needle tip puncture tester, according to GB 15811-2016 disposable sterile injection needle standard in the injection needle puncture force test and evaluation method, the principle of the injection needle puncture force test is to use the puncture force test device to make the injection needle at a specified speed, vertically through the maximum peak measured when simulating the skin to evaluate the puncture force
    of the injection needle.

    The exact amount of "prescribed speed" will be mentioned
    in the test procedure of the needle tip puncture force tester.

    The test procedure of the needle tip puncture force tester is as follows: 1.
    The needle and the simulated skin will be placed at 22 ° C at least 2 ° C for at least 2 4 h, and tested
    at the same temperature.

    2.
    According to the measuring device shown in Figure B.
    1, the simulated skin of appropriate size shall be clamped on the fixture, and no obvious stretching or compressive force shall be applied to the
    simulated skin.

    3.
    The needle will be mounted on the test device (B.
    2.
    1), its axis is perpendicular to the surface of the simulated skin (B.
    2.
    3), and the needle tip points to the center
    of the circular puncture area.

    4.
    Set the movement speed of the test device (B.
    2.
    1) to 100 mm/min
    .

    5.
    Start the test device
    .

    6.
    In the process of puncture on the membrane, the maximum peak force or recorded force/displacement diagram is measured at the
    same time.

    (Note: Do not use membranes
    that have been punctured in circular puncture areas.
    )
    How much force is used during the puncture? The recommended force value is as follows: the injection needle with a specification of 0.
    3~0.
    6mm should use a force value of less than or equal to 0.
    7N during the test; Injection needles with a size of 0.
    9~0.
    9mm use a force value of less than or equal to 0.
    85N during the test; Injection needles with a size of 1.
    1 to 1.
    2 mm use a force value of less than or equal to 1.
    15 N
    during the test.

    Warm reminder In the injection needle physical performance test, experienced operators know that the injection needle tip piercing tester, injection needle tip rigidity tester, injection needle tip toughness tester, injection needle tip fracture force tester to detect the piercing force, rigidity, toughness, breaking force and other properties of the needle tip, but don't forget to use the injection needle to connect the firmness tester and 6% The Ruhr conical connector multifunctional tester checks the firmness of the connection between the needle holder and the needle cap and the performance
    of the ruhr on the injection needle spare part.

    The above content is provided by Viscia Technology, I hope to help
    you.

    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.